Automated detection of organophosphate warfare gases (nerve agents) in air based on micro-SIA - lab-on-valve system
Ondrej Pavlicek, Miroslav Polasek, Martin Foltyn, Jiri Cabal
Address: Jiri Cabal, Faculty of Military Health Sciences, University of Defense, POB 35-T, 500 01 Hradec Kralove, Czech Republic
cabal@pmfhk.cz
Received 24
th February 2012.
Revised 2
nd April 2012.
Published online 6
th April 2012.
Full text article (pdf)Full text article (html)
SUMMARY
Equipment for fast and accurate detection of organophosphate nerve agents is developed and tested. The method is based on the spectrophotometric
monitoring of the enzyme activity of butyrylcholinesterase after its contact with air in a special absorption unit (a “scrubber”) developed for the
purpose. The scrubber was made from a glass tube filled with glass beads (diam. 3 mm) and filled with approx. 5 ml of butyrylcholinesterase in a
phosphate buffer of pH 7.4. The air sample was bubbled through this solution for 20 s at a flow rate of 80 l hour-1. Thereafter 8 microl of
the enzyme solution were aspirated into the micro-SIA-LOV analyzer and the activity of the enzymes were evaluated by using Ellman’s reagent, i.e. 2.5
mmol l-1 butyrylthiocholine iodide and 0.25 mmol 5,5’-dithiobis (2-nitrobenzoic acid). The absorbance of the coloured reaction product was
measured at 412 nm after the reaction time of 60 s. The residue of the absorption liquid was washed away from the absorber and the system was washed
with the enzyme solution prior to next analysis. The contaminated air caused partial inhibition of the enzyme activity of the absorption liquid. The
activity of the contaminated sample was compared with the activity of the unaffected enzyme (blank measurement). The analysis was controlled by two
PCs. The effect of the concentration of analyte in the absorption liquid on the enzyme activity was tested for 10-5-10-9 mol
l-1 sarin. A single analysis (including the absorption step) took <130 s.
KEY WORDS
sequential injection analysis; lab-on-valve; organophosphate; cholinesterase; nerve agent; sarin
REFERENCES
Ellman GL, Courtney DK, Andres V. A new and rapid colorimetric determination of acetylcholinesterase activity. Biochem Pharmacol. 7: 88-90, 1961. [CrossRef]
Maruko H, Sekiguchi H, Seto Y, Sato A. Detection performance of aspiration-type ion mobility spectrometer for chemical warfare agents. Bunseki Kagaku.
55: 191-197, 2006. [CrossRef]
Matsushita K, Sekiguchi H, Seto Y. Performance of portable surface acoustic wave sensor array detector for chemical agents (in Japanese). Bunseki
Kagaku. 54: 83-88, 2005. [CrossRef]
Mesillakso M. Chemical Weapons Convention Chemicals Analysis. Sample Collection, Preparation and Analytical Methods. J. Wiley & Sons, Chichester, 2005,
pp. 67-77. [CrossRef]
OPCW. Convention on the Prohibition, Stockpiling and Use of Chemical Weapons and on their Destruction. Organisation for the Prohibition of Chemical
Weapons, The Hague, 1994.
Riegle DW, D'Amato AM. U. S. Chemical and Biological Warfare-Related Dual Use Exports to Iraq and their Possible Impact on the Health Consequences of
the Gulf War. United States Senate, 103rd Congress, 2nd Session, May 25, 1994, http://www.gulfweb.org/bigdoc/report/appgb.html,
downloaded: 7th Feb 2012.
Ruzicka J. Lab-on-valve: Universal microflow analyzer based on sequential and bead injection. Analyst. 125: 1053-1060, 2000. [CrossRef]
Sevelova L, Vachek J, Bajgar J. Inhalation apparatus for generating sarin and soman toxic vapors. Acta Medica (Hradec Kralove). 46: 157-161, 2003. [PubMed]
|
CITED
0
|
BACK